4. Safety functions
4.1. Emergency stop
On the left hand side, you will find the emergency stop (See
figure 13). Operate this button in an emergency; if the upward
or downward movement must be stopped immediately . Once
the emergency stop has been pressed, the electronic controls
will stop working . Operation may be continued once the
situation is stable again . Turn the emergency stop a quarter
turn to unlock it again (see Figure 14).
Check whether the danger has passed or the problem
has been resolved before resetting the emergency stop.
The orange LED light on the hand control will flash
when the emergency stop has been activated.
4.2. Sleep setting
After around an hour, the chair will fall into a 'sleep' setting
and the battery indication LEDs on the handcontrol will go
out . In order to use the chair again, it must be activated
(See paragraph 2.2 Activating).
4.3. Emergency lowering
If the controls stop working and the chair is in a raised posi-
tion, it can be lowered using the emergency lowering button .
Check first to make sure the chair is not in sleep setting!
• Remove the plug from the handcontrol (1) (See figure 15) .
• Press the emergency lowering button (2) (See figure 15) .
The chair will now return to the lowest position (make sure no persons or objects are
under the chair) .
Take the client from the chair and make sure the device is not used until it has been
repaired . Place the chair in a safe place and indicate that it is out of order .
Contact the technical service or Lopital's service department . E-mail: service@lopital .nl
34 | Instructions for use | Elexo XXL Shower-toilet chair
Figure 13 Emergency stop .
Figure 14 Emergency stop .
2
1
Figure 15
Emergency lowering button .