Immunity test
Electrical fast
transient/ burst
IEC 61000-4-4
Surge IEC
61000-4-5
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
•
English
38
IEC 60601
Compliance
Test level
level
�2 kV for
power
supply
lines
Not
�1 kV for
applicable
input/
output
lines
�1 kV
differential
mode
Not
�2 kV
applicable
common
mode
3A/m
Complies
> 95% dip
Not
0.5 cycle
applicable
60% dip
5 cycles
70% dip
25 cycles
95 dip 5
sec.
Electromagnetic
environment - guidance
The ME equipment is
solely battery powered.
Power frequency
magnetic fields should
be at levels characteristic
of a typical location in a
typical commercial or
hospital environment.
The ME equipment is
solely battery powered.