Important information regarding

14

electromagnetic compatibility (EMC)

MEDICAL ELECTRICAL EQUIPMENT needs special precautions
regarding EMC. Portable and mobile RF communications
equipment can affect MEDICAL ELECTRICAL EQUIPMENT.
Non-life support equipment separation distance calculation
Rated maximum
output power
of transmitter
(Watt)
0.01
0.1
1
10
100
English
36
Please check if you have hold
the thermometer in your
hand too long and affect the
accuracy
Please check if there is low
battery icon on the LCD
display
(3Vrms/3V/m compliance)
Separation distance according to frequency of
150kHz to
80Mhz
d = 1.2 √P
0.12
0.37
1.17
3.69
11.67
Put the thermometer
on the table in the room
where the measurement
is taking place and let it
cool down first.
Replace with new
batteries
transmitter (meter)
80Mhz to
800Mhz
d = 1.2 √P
0.12
0.37
1.17
3.69
11.67
800Mhz to
2.5Ghz
d = 1.2 √P
0.23
0.74
2.33
7.38
23.33