4.5
Emergency Stop
IMPORTANT!
When the emergency stop is activated, the Reset
function can still be activated, by pressing button A,
to lower the wings to horizontal position if required.
When the emergency stop is activated it stops all controls and
movements immediately (wings will stay in current position).
1.
To activate, push the button .
2.
To deactivate, rotate the button clockwise.
3.
Press button A to reset the system.
4.6
Manual operation
WARNING!
Risk of squeezing and injury
If both wings are simultaneously elevated to angles of
more than 12° there is a risk of squeezing the patient.
–
This option must only be used by trained personnel.
The first time MANUAL is activated, after the system has
been switched on, it will reset itself before the function
can be used.
1.
Press and hold button for 3 sec. to activate manual
operation.
2.
Press buttons on hand control as required.
•
To raise the left wing, press
•
To lower the left wing, press J
•
To raise the right wing, press K
•
To lower the right wing, press L
•
To raise right wing from 0
°
5
, press N
•
To raise left wing from 0
°
5
, press M
- Left and right buttons can be activated at the same time
to move both wings simultaneously. The wings can be
moved in a range from 0° to 80°.
3.
To deactivate manual operation, press the ON/OFF button
A.
IVCMAMED221119
°
to 80
°
and left wing from 0
°
°
to 80
and right wing from 0
Status indicators
Display
0 = no activity
Uxx = right wing
moves up
dxx = right wing
moves down
xxU = left wing
moves up
xxd = left wing
moves down
4.6.1 Recommended Procedures
Transfer from flat position to the side:
Press either N or M depending on which side the patient
should be elevated to.
Transfer from one side to the other:
1.
If left side is elevated - press and hold N to transfer from
left to right side. Release button when desired height is
maintained.
2.
If right side is elevated - press and hold M to transfer from
right to left side. Release button when desired height is
maintained.
4.7
Automatic operation
The AUTO program is pre-defined and can not be changed.
Program 1 and 2 can be individually defined by the care staff (see
chapter Programming). Upon delivery the programs are empty
and need to be defined before usage.
- If one of the programs is activated, the system will
reset to FLAT position before starting the selected
program.
- If a new program is selected while another program
is active, the system will reset to FLAT position before
starting the new program.
°
to
°
to
1.
Press program buttons as required.
•
To start the AUTO Program, press and hold O for 3 sec.
•
To start Program 1, press and hold P for 3 sec.
•
To start Program 2, press and hold Q for 3 sec.
LED
Note
MANUAL LED
The 'xx' letters on the
active
display represent the
angle at which the wing is
positioned, while the letters
represent the direction of
movement:
- U = up movement.
- d = down movement.
(degrees)
- xx =x
Example: 10=10 degrees.
or
or
Usage
11