Immunity test levels
Test
Electrostatic Discharge
(IEC 61000-4-2)
Radiated RF EM fields
(IEC 61000-4-3)
Electrical fast transients / bursts
(IEC 61000-4-4)*
Surges
(IEC 61000-4-5)*
Conducted disturbances
induced by RF fields
(IEC 61000-4-6)*
Proximity fields from RF wireless
communications equipment
(IEC 61000-4-3)
Power frequency magnetic fields
(IEC 61000-4-8)
Voltage dips
(IEC 61000-4-11)*
Short interruptions
(IEC 61000-4-11)*
*n/a: "Not applicable" in the internally powered mode
IEC 60601-1-2 test levels
Contact Discharge: ±8 kV
Air Discharge: ±2 kV, ±4 kV, ±8 kV, ±15 kV
3 V/m
80 – 2700 MHz
80 % AM at 1 kHz
± 2 kV
100 kHz repetition frequency
± 0.5 kV, ± 1 kV
3 V
0.15 MHz – 80 MHz
6 V in ISM bands between 0.15 MHz and 80 MHz
80 % AM at 1 kHz
385 MHz; Pulse Modulation: 18 Hz; 27 V/m
450 MHz, FM: ±5 Hz deviation: 1 kHz sine; 28 V/m
710, 745, 780 MHz; Pulse Modulation: 217 Hz; 9 V/m
810, 870, 930 MHz; Pulse Modulation: 18 Hz; 28 V/m
1720, 1845, 1970 MHz; Pulse Modulation: 217 Hz; 28 V/m
2450 MHz; Pulse Modulation: 217 Hz; 28 V/m;
5240, 5500, 5785 MHz; Pulse Modulation: 217 Hz; 9 V/m
30 A/m; 50Hz or 60 Hz
0 % U T ; 0.5 cycle; at 0°, 45°, 90°,
135°, 180°, 225°, 270° and 315°
0 % U T ; 1 cycle and 70 % U T ;
25/30 cycles Single phase: at 0°
0 % U T ; 250/300 cycles
Test level
Compliance test levels
DELTA 30/DELTA 30 PRO:
Compliance test levels are
identically to the IEC 60601-1-2
test levels
DELTAone: Test was not
necessary, because there are no
magnetic sensitive components.
DELTA 30/DELTA 30 PRO:
Compliance test levels are
identically to the IEC 60601-1-2
test levels
DELTAone: Test was not
performed, because it was
not considered necessary.
No risks were identified.
med 236321 2023-09-07 59/68