NUMBER OF CHANNELS
GB
CONSTANT CURRENT
STIMULATION FORMS
AMPLITUDE
WAVEFORM
MAXIMUM FREQUENCY
MAXIMUM PULSE DURATION
POWER SOURCE
CURRENT CONSUMPTION FOR ONE
CHANNEL, 180 µS, 80 Hz, 30 mA
ENVIRONMENT FOR
STORAGE, USE
AND SHIPPING
EXTERNAL DIMENSIONS
WEIGHT
R.M.S. MAX/CHANNEL
Warranty: The stimulator is covered by a two-year warranty against manufacturing faults.
Subject to normal conditions of use.
10
All manuals and user guides at all-guides.com
1
Up to a resistance of 1000 Ω. (increased load can
reduce the maximum current)
Continuous stimulation: modulated pulse duration,
modulated frequency.
0-60 mA, in steps of 1 mA (charge maximum: 18μC)
Symmetrical biphasic square pulse
80 Hz
300 µs
2 x 1.5 V AA non-rechargeable or
2 x 1.2 V AA rechargeable batteries
80 mA
Temperature 10º C - 40º C
Air humidity 30% - 75%
Air pressure 700 hPa-1060 hPa
52 x 125 x 30 mm
approximately 130 g incl. batteries
10 mA