Replacing or refilling containers

The DOS modules are equipped with a monitor which sends a mes-
sage to the washer-disinfector if the level drops below the minimum.
Fill the relevant container with chemical agent when the message
appears in the washer-disinfector's display.
Unscrew and remove the siphon and place it on the drip tray.
Fill or replace the container, fit the siphon and secure the lid.
Feed the siphon into the container until it reaches the bottom.
Make sure the hose is not kinked.
Refill the container in good time to prevent the dispensing system
being emptied completely.
If a container has been emptied, the dispensing system has to be
vented before the start of the next programme.
Follow the information in the operating instructions for the washer-
disinfector.
The drip trays located underneath the containers can be removed
for cleaning and can be cleaned in the washer-disinfector.
Clean the drip trays from time to time.

Maintenance

Maintenance of the dispensing system can be performed by Miele
Customer Service at the same time as maintenance of the cleaning
machine.