3H
Check whether the two outlets draw in air. In order to check the performance of the pump,
you should have a spare breast pump kit to be able to assemble the complete pump.
How to check the performance of the pump:
-
Assemble all parts of the breast pump kit: silicone tube and cap, pump body with white
valve, Natural bottle and diaphragm (Fig. B).
-
Press and hold the on/off button (A2 in Fig. A) until the device switches on and press
the 'Level up' button (A5 in Fig. A) several times.
-
Check if the silicone diaphragm (B2 in Fig. B) moves gradually to the top of the cap and
all the way back down in each pumping cycle.
-
The above-described checks must be performed for both diaphragms.
-
If any of diaphragms does not move according to the description above, please contact
your Philips sales agent.
-
3I
Check the travel bag and insulation pouch to make sure they are not damaged.
-
Check the user manual for completeness.
-
If the travel bag, insulation pouch or the user manual is damaged, please replace.

Step 4: Factory reset

Reset the motor unit to factory settings as follows:
4A. Make sure that the
adapter is inserted into
the wall socket and that
the small plug at the
other end is inserted
into the motor unit.

Step 5: Cleaning before renting

5A
Motor unit
Adapter
5B
Travel bag
Insulation
pouch
5C
User manual
3 sec.
4B. Press and hold the
on/off button to switch
off.
It is important to repeat cleaning of the motor unit (the buttons and overall
surface) and the adapter before the breast pump is rented out to the next
customer. This can be done by using alcohol wipes (isopropanol).
Clean the travel bag and the insulation pouch by wiping them with a clean,
damp cloth with water and mild detergent.
Warning: Do not use alcohol wipes (isopropanol) to clean the
travel bag and the insulation pouch.
Do not clean the user manual. If the user manual is dirty or damaged, please
replace it.
4C. Simultaneously
press the mode
selection button (see
A3 in Fig. A) and the
'Level up' button (see
A5 in Fig. A).
4D. Wait until the
indicator lights for
modes and levels light
up (A6 in Fig. A).

Advertenties

Inhoudsopgave
loading

Deze handleiding is ook geschikt voor:

Avent scf371/11Avent scf338/11

Inhoudsopgave