15 Manufacturer's EMC guidelines
Electromagnetic interference emissions
The Tel-O-Graph
®
BT is intended to be operated in the electromagnetic environment specified below. Only use the
Tel-O-Graph
®
BT in such an environment.
Measurements of
interference
emmisions
RF interference
emissions according to
CISPR 11
RF interference
emissions according to
CISPR 11
RF interference
emissions according to
CISPR 25
IEC 61000-3-2
IEC 61000-3-3
Compliance
Group 1
Class B
Not applicable
Not applicable
Not applicable
Manufacturer's EMC guidelines
Electromagnetic environment
The Tel-O-Graph
®
BT uses RF power exclusively for
its internal function. Its RF emission is therefore
very low and it is unlikely that any neighbouring
electronic device will experience any interference.
The Tel-O-Graph
®
BT is suitable for use in facilities
other than residential areas and those directly
connected to the public supply network which also
supplies buildings used for residential purposes.