TECHNICAL FEATURES
Mod. P0404EM F400
Voltage:
Safety certifications:
Max pressure:
Compressor air output:
Sound level (at 1 m):
Operation:
Operating conditions:
Temperature:
RH Air humidity:
Atmospheric pressure:
Storage conditions:
Temperature:
RH Air humidity:
Atmospheric pressure:
Dimensions (W)x(D)x(H):
Weight:
APPLIED PARTS
Type BF applied parts are:
RF6 Basic
nebuliser
2
Medication minimum capacity:
Medication maximum capacity:
Operating pressure (with neb.):
MMAD (μm)
Standard
High speed
mode
3.8
Data detected according to Flaem I29-P07.5 internal procedure
(1)
In vitro testing certified by TÜV Rheinland LGA Products GmbH - Germany in compliance with the European
(2)
Standard EN 13544-1 for nebulising systems. Further details are available upon request.
DISPOSAL OF DEVICE:
In conformity with Directive 2012/19/EC, the symbol shown on the device to be
disposed of indicates that it is considered as waste and is therefore subject to
"sorted waste collection". The user must therefore take (or have taken) the above
waste to a pre-sorted waste collection centre set up by the local authorities, or
else give it back to the dealer when purchasing a new appliance of the same
type. Pre-sorted waste collection and the subsequent treatment, recovery and disposal
operations favor the production of appliances made of recycled materials and Iimit the
negative effects of any incorrect waste management on the environment and public
health. The unlawful disposal of the product by the user could result in administrative
fines as provided by the laws transposing Directive 2012/19/EC of the European
member state or of the country in which the product is disposed of.
230V~ 50Hz 130VA
BREATHABLE
FRACTION < 5 μm
(2)
Standard
mode
mode
4.0
62%
115V ~ 60Hz
1.8 ± 0.3 bar
9 l/min approx
56 dB (A) approx
Continuous
min 10°C; max 40°C
min 10%; max 95%
min 69KPa; max 106KPa
min -25°C; max 70°C
min 10%; max 95%
min 69KPa; max 106KPa
16x16.5x11.5 cm
1.300 Kg
patient accessories (C2, C3, C4)
2 ml
8 ml
0.6 bar
(FPF)
(2)
High speed
mode
63%
8
220V ~ 60Hz
DELIVERY ml/min
(approx)
Standard
High speed
mode
mode
0.25
0.32
(1)