ENG
Clean the lens with a nonabrasive cleaning cloth (such
as an eyeglass cloth) and a suitable cleaning agent (see
"Recommended disinfectants").
CAUTION
To minimize the risk of disease transmission, in
addition to complying with this operating manual, you
must also comply with the applicable occupational
health and safety regulations and the requirements of
authorities responsible for hygiene and disinfection.
6

SAFETY INSPECTIONS

DANGER
Death hazard from electric shock
Unplug the charger and turn off the luminaire.
The charger and charging cable must be checked at
least once a year for damage.
CAUTION
Maintenance and repairs can only be performed by
qualified technical personnel.
The corresponding user profile is in Section 1 Safety
instructions.
EVERY YEAR
Check the charger and charging cable for damage
and replace if necessary.
Check for deformations and cracks in the plastic parts.
Check for loose parts.
9.

TROUBLESHOOTING

Fault
The luminaire doesn't go on
Luminaire can't be charged
Luminaire can't be charged
Luminaire doesn't go on, red
LED flashes
18
Possible cause
Dead battery
No power supply
Charger / charging cable
defective
Defective electronic
component
7.

REPAIRS

DANGER
Death hazard from electric shock and fire
For service, the luminaire must be unplugged from the
power supply before removal.
Only original parts are to be used for all repairs.
Damaged batteries must never be reused.
7.1

Disposal

Do not discard the luminaire with household
waste. Follow local regulations and take the
luminaire to a disposal site or hand it over to a
dealer with an appropriate service department.
The products listed above are over 95%
recyclable. For a high percentage of the used
materials to either be physically reused or used
for energy after the end of their life cycle, the
luminaires have been designed with recycling in
mind.
The battery contains materials subject to
monitoring.
The battery must be separated from the rest of
the
product
according to national regulations.
Before disposal, the battery must be completely
drained and the contacts must be insulated.
8

ADDITIONAL INFORMATION

The luminaire itself is maintenance free.
Additional
documents
manufacturer for this product.
Using this luminaire does not pose a risk to other equipment.
To save energy, the luminaire should be switched on only
when it is actually needed.
Any serious incident that has occurred with the product must
be reported to the manufacturer or their representative and
the responsible authorities of the member state in which the
user is located.
Correction
Fully charge the battery
Check voltage, check all connections
Replace charger / charging cable
Contact manufacturer's service
department
and
disposed
of
separately
may
be
requested
from
User profile
All
Qualified electrician
All
Only by manufacturer's
service department
the