17
Technical Information
RS 232/RS 485 Communication Description
N.
Read/ Write
202
R
203
R
204
R
205
R
206
R
210
R (see note 3.)
300
R
301
R
302
R
320 to
Reserved to Agilent service
399
400
R
402
R
404
R
196/208
Data
Description
Type
N
Pump power in W (pump current
x pump voltage duty cycle)
N
Driving frequency in Hz
N
Pump temperature in °C
N
Pump status
N
Error code
N
Actual rotation speed in Hz
N
Cycle time in minutes (zeroed by
the reset command)
N
Cycle number (zeroed by the
reset command)
N
Pump life in hours (zeroed by the
reset command)
A
CRC EPROM (QE)
A
CRC Param. (PA)
A
CRC Parameter structure
Special TV 301 Navigator Controller User Manual / 87-900-981-01 (E)
Admitted
Values
0 to 70
Stop = 0
Waiting intlk = 1
Starting = 2
Auto-tuning = 3
Braking = 4
Normal = 5
Fail = 6
Bit description: see
the following figure
100 to 963
0 to 999999
0 to 9999
0 to 999999
QE8XXXX
(where "XXXX" are
variable)
PA8XXXX
(where "XXXX" are
variable)
"XXXX"