17
Technical Information
RS 232/RS 485 Communication Description
Window Meanings
Tab. 5
N.
Read/ Write
000
R/W
008
R/W
100
R/W
101
R/W
102
R/W
103
R/W
104
R/W
105
R/W
106
R/W
107
R/W
194/208
Data
Description
Type
L
Start/Stop
(in remote mode the window is a
read only)
L
Remote
(default) or Serial configuration
L
Soft Start
(write only in Stop condition)
N
Set Point type
N
Set Point
threshold
(expressed in Hz, mA or s)
N
Set Point mask: time between
the pump start and the set point
check (seconds)
L
Set Point signal activation type:
the signal can be "high level
active" or "low level active"
N
Set point
hysteresis
(in % of threshold)
L
Water cooling
L
Active Stop
(write only in stop)
Special TV 301 Navigator Controller User Manual / 87-900-981-01 (E)
Admitted
Values
Start = 1
Stop = 0
Remote = 1
Serial = 0
(default = 1)
YES = 1
NO = 0
0 = Frequency
1 = Current
2 = Time
(default = 0)
(default = 850)
0 to 99999
(default = 13.800 (46
min.))
0 = high level
active (N.O.)
1 = low level active
(default = 0)
0 to 100
(default = 12)
0 = NO
1 = YES
0 = NO
1 = YES