N.
Read/ Write
108
R/W
109
W
110
R/W
111
R/W
120
R/W
121
R/W
122
R/W
123
Reserved to Agilent service
124
125
R/W
126
R/W
130
Reserved to Agilent service
200
R
201
R
Special TV 301 Navigator Controller User Manual / 87-900-981-01
RS 232/RS 485 Communication Description
Data
Description
Type
N
Baud rate
L
Pump life/ cycle time/ cycle
number reset
L
Interlock type
(default = 1)
L
Analog output type: output
voltage signal proportional to
frequency or power
N
Rotational frequency setting (Hz)
N
Maximum rotational frequency
in Hz (active only in Stop
condition)
L
Set vent valve on/off
(on = closed)
L
Set the vent valve
operation
N
Vent valve opening
delay (expressed in 0.2 sec)
N
Pump current in mA dc
N
Pump voltage in Vdc
Technical Information
Admitted
Values
600 = 0
1200 = 1
2400 = 2
4800 = 3
9600 = 4
(default = 4)
To reset write '1'
Impulse = 0
Continuous = 1
0 = frequency
1 = power
(default = 0)
150 to 963
(default = 963)
150 to 963
(default = 963)
On = 1
Off = 0
(default = 1)
Automatic = 0 (see
note 1.)
On command = 1
(see note 2.)
0 to 65535
(corresponding to 0
to 13107 sec)
(E)
195/208
17