W220 Peristaltic pump Thank you very much for purchasing our product,please keep and read this manual carefully before you install the Peristaltic pump _________________________________________________________________________...
Packing List The W'eau W220 pump is supplied complete with: ● Bracket for wall installation ● Standard kit including suction and head tube, foot filter, and injection valve ● Power cable 1.5 m (internally pre-wired) ● Instruction manual Technical Data ●...
Installation Install the W'eau W220 pump in a dry area at a maximum ambient temperature of 45°C and place it in a way that allows easy adjustment and maintenance operations. Wall mount the pump using the supplied bracket. Secure the bracket to a vertical wall (±15°) and attach the pump.
Pagina 4
2. Cut the PVC Crystal tube and insert the tube fixing nut onto the tube. 3. Mount the tube on the conic hose of the suction fitting, pushing it to reach the stop collar. 4. Secure the tube by screwing the tube fixing nut on the suction fitting of the pump body.
Electrical Connections Power supply: 230 V~ 50/60 Hz.The mains circuit must be protected in accordance with the relevant laws and regulations. Typically, the protection is given by a 30 mA differential switch and a breaker or 1 A fuse. ● If a level control sensor is used, plug it to its connector (see figure). Warning! If several pumps are connected in parallel, always observe the polarity of the level connection to not endanger the proper functioning of the system or damage the inputs!
Pagina 6
● ON/OFF key: enables/disables the system; press and hold for at least 3 seconds to enter the CONFIGURATION mode. ● [-] key: press and hold to display the electrode offset value (for 3 seconds) and then enter the OFFSET CALIBRATION mode. ●...
Pagina 7
Storage of the ON/OFF key status at shutdown: 0 = no, 1 = yes Control type: 0 = OFF, 1 = ON, 2 = ON/OFF acid, 3 = Proportional acid, 4 = ON/OFF chlorine, 5 = Proportional chlorine Working threshold: 0 to 14.0 (pH), 0 to 999 (mV) 7.00 Hysteresis: 0 to 2.00 (pH), 0 to 200 (mV) 0.40...
Pagina 8
● P3: This parameter allows storing the status of the ON/OFF button at the system shutdown: 0 = at start-up, the pump is always enabled; 1 = at start-up, the pump restores its status at shutdown. ● P4: This parameter allows setting the pump working mode: ●...
Pagina 9
Pump OFF always 67 s 45 s 23 s never ● 4 = ON/OFF chlorine (this option is typically used for chlorination or alkalization; the pump activates when the measurement is lower than the “working threshold – ½ hysteresis” value and deactivates when the measurement exceeds the “working threshold + ½...
Pagina 10
pH (or between zero and 200 mV), while in the case of proportional controls, the hysteresis must have a value between 0.20 and 1.00 pH (or between 20 and 100 mV). ● P7: At the device start-up, some electrodes require a stabilization (or polarization) time during which the measurement is not reliable.
Pagina 11
5. Rinse the electrode with distilled water and then immerse it in the pH 4.01 (or 9.01) buffer solution. 6. Wait a few seconds for the system stabilization. 7. Press and hold the [+] key until the message “GAI” (GAIN calibration) is displayed.
Pagina 12
Regular maintenance is essential for the pump to give good service over a long period. The following advice should be strictly followed. Before any operation, make sure the system is unplugged! Weekly operations: ● Check the level of the liquid to be dosed to avoid the pump working dry. ●...
Pagina 13
Extraordinary maintenance – Replacing the peristaltic tube: 1. Unscrew the nuts and remove the suction and injection hoses, then remove the transparent cover by unscrewing the two front screws. 2. Remove the old tube first by unlocking the left fitting and turning the roller-holder as indicated by the arrow in the figure to release the tube from the fitting at the right.
Pagina 14
● Check that the foot filter is not clogged. ● Check that the injection valve is not clogged. ● Check the suction and head valves. Liquid leakages from pump body: ● Check that the tubes are well inserted and the nuts well tightened. ●...
INSTALLATIE Installeer de W'eau W220-pomp in een droge omgeving bij een maximale omgevingstemperatuur van 45°C en plaats deze zo dat eenvoudige afstel- en onderhoudswerkzaamheden mogelijk zijn. Monteer de pomp aan de muur met de meegeleverde beugel.
Pagina 17
Opmerking: Plaats de voetfilter op een minimale afstand van 5 cm van de bodem van de tank. Als er een stroperig product wordt gedoseerd, is het raadzaam het interne filter te verwijderen om de aanzuiging te vergemakkelijken. Injectieleiding (zie ook tekening) 1.
● Sluit de meetsensor aan op de BNC-connector. BEDIENINGSPANEEL ● Display: toont tijdens normale werking de gemeten pH of redox (mV); de volgende meldingen kunnen ook verschijnen: ● OFF: de pomp is uitgeschakeld door op de ON/OFF-knop te drukken. ● LEU: de niveausensor geeft geen toestemming. ●...
Pagina 19
Opslag van de status van de ON/OFF-knop bij uitschakeling: 0 = nee, 1 = ja Regeltype: 0 = UIT, 1 = AAN, 2 = AAN/UIT zuur, 3 = Proportioneel zuur, 4 = AAN/UIT chloor, 5 = Proportioneel chloor Werkdrempel: 0 tot 14,0 (pH), 0 tot 999 (mV) 7,00 Hysterese: 0 tot 2,00 (pH), 0 tot 200 (mV) 0,40...
Pagina 20
hysterese” overschrijdt en gedeactiveerd wanneer de meting onder de “werkdrempel – ½ hysterese” komt). ● 3 = Proportioneel zuur (de pomp is zeker actief wanneer de meting groter is dan de “werkdrempel + ½ hysterese” en zeker uit wanneer de meting lager is dan de “werkdrempel –...
● P5: Deze parameter maakt het mogelijk de werkdrempel in te stellen die wordt gebruikt wanneer de pomp automatisch werkt. ● P6: Deze parameter vertegenwoordigt de interventiehysterese rond de werkdrempel; bij ON/OFF-regelingen kan deze worden ingesteld tussen nul en 2,00 pH (of tussen nul en 200 mV), terwijl bij proportionele regelingen de hysterese een waarde moet hebben tussen 0,20 en 1,00 pH (of tussen 20 en 100 mV).
Pagina 22
3. Houd de [-] knop ingedrukt totdat het bericht "OFS" (OFFSET-kalibratie) wordt weergegeven. 4. Druk op ON/OFF om de kalibratie te bevestigen of wacht een paar seconden om te verlaten zonder op te slaan en de vorige kalibratie te behouden. 5.
ONDERHOUD Regelmatig onderhoud is essentieel om de pomp langdurig goed te laten functioneren. Volg het onderstaande advies strikt op. Voordat u enige handeling uitvoert, zorg ervoor dat het systeem is losgekoppeld! Wekelijkse handelingen: ● Controleer het niveau van de te doseren vloeistof om te voorkomen dat de pomp droog werkt.
2. Schroef de 6 schroeven los die de behuizing vasthouden en open het achterpaneel. 3. Vervang de doorgebrande zekering op het circuit door een nieuwe van hetzelfde type (afmetingen en afmetingen). 4. Als de zekering opnieuw doorbrandt, stuur de pomp dan naar de fabrikant voor reparatie.
Pagina 25
● Controleer het display en de configuratie: de pomp kan zijn uitgeschakeld of een alarm is actief. De pomp werkt correct maar injecteert geen vloeistof in de installatie: ● Controleer het productniveau in de tank. ● Controleer of het voetfilter niet verstopt is. ●...