10.6 Calibrating the scale
The scale is pre-calibrated before delivery. If the measurements are inaccurate, try
replacing the batteries first. If the accuracy does not improve, you can recalibrate the
scale using a suitable calibration weight of 200 g or 500 g.
1. Press and hold the CAL button (8) until the calibration weight (200 g or 500 g)
is displayed.
2. Press the TARE button (5) until the desired calibration weight (200 g or 500 g)
is displayed.
3. Press the CAL button (8) again.
"CAL" will appear on the display and the selected calibration weight will
Æ
flash.
4. Place the calibration weight on the weighing plate (1).
5. Wait until "PASS" appears on the display.
6. Remove the calibration weight from the weighing plate (1).
Calibration is now complete.
Æ
10.7 Setting the power-off timer
By default, the scale powers off after 5 minutes of inactivity. Alternatively, you can
set a different power-off time:
1. Press and hold the TARE button (5) until the power-off time appears on the
display.
2. Press the TARE button (5) several times until the desired power-off time is
reached:
– "1" (1 min) → "3" (3 min) → "5" (5 min) → "-" (disabled)
3. Press the MODE button (4) to confirm your changes.
The power-off time is now set.
Æ