PV, SV, TV en QV kunnen via de sensor-DTM worden ingesteld.
Register Name
Register Number
Status
100
PV Unit
104
PV
106
SV Unit
108
SV
110
TV Unit
112
TV
114
QV Unit
116
QV
118
Status
1300
PV
1302
SV
1304
TV
1306
QV
1308
Status
1400
PV
1402
Status
1412
SV
1414
Status
1424
TV
1426
Status
1436
QV
1438
Status
2000
PV
2002
SV
2004
TV
2006
QV
2008
Status
2100
PV
2102
SV
2104
VEGADIF 85 • Modbus- en Levelmaster-protocol
Type
Note
DWord
Bit 0: Invalid Measurement Value PV
Bit 1: Invalid Measurement Value SV
Bit 2: Invalid Measurement Value TV
Bit 3: Invalid Measurement Value QV
DWord
Unit Code
Primary Variable in Byte Order CDAB
DWord
Unit Code
Secondary Variable in Byte Order CDAB
DWord
Unit Code
Third Variable in Byte Order CDAB
DWord
Unit Code
Quarternary Variable in Byte Order CDAB
DWord
See Register 100
Primary Variable in Byte Order of Register 3000
Secondary Variable in Byte Order of Register 3000
Third Variable in Byte Order of Register 3000
Quarternary Variable in Byte Order of Register 3000
DWord
See Register 100
Primary Variable in Byte Order CDAB
DWord
See Register 100
Secondary Variable in Byte Order CDAB
DWord
See Register 100
Third Variable in Byte Order CDAB
DWord
See Register 100
Quarternary Variable in Byte Order CDAB
DWord
See Register 100
DWord
Primary Variable in Byte Order ABCD (Big Endian)
DWord
Secondary Variable in Byte Order ABCD (Big Endian)
DWord
Third Variable in Byte Order ABCD (Big Endian)
DWord
Quarternary Variable in Byte Order ABCD (Big Endian)
DWord
See Register 100
DWord
Primary Variable in Byte Order DCBA (Little Endian)
DWord
Secondary Variable in Byte Order DCBA (Little Endian)
11 Bijlage