Data
(1) Activate limitation
measuring range begin
(2) Manual limitation of the
measuring range begin
(5) Averaging factor on in-
creasing amplitude
(6) Averaging factor on de-
creasing amplitude
(15) Activate measure-
ment of the "first large echo"
function
(16) Amplitude difference
"First large echo" function
(25) Adjustment in
Distance
Measurement reliability
Filling height
Electronics temperature
Measure rate
Switching output 1
Switching output 2
Current output
Device status acc. to
NE 107
Device status
Detail status
Counter for change of pa-
rameters
Filling height
Distance
Measurement reliability
Electronics temperature
Meas. rate
Switching output 1
Switching output 2
Current output
Function control
Out of specification
Maintenance required
VEGAPULS 42 • Driedraads met IO-Link (2 x transistor of 4 ... 20 mA plus 1 x transistor)
ISDU
ISDU
(dez)
(hex)
341
0x0155
342
0x0156
343
0x0157
344
0x0158
345
0x0159
346
0x015A
347
0x015B
348
0x015C
349
0x015D
350
0x015E
351
0x015F
352
0x0160
353
0x0161
354
0x0162
355
0x0163
356
0x0164
357
0x0165
358
0x0166
359
0x0167
360
0x0168
361
0x0169
362
0x016A
363
0x016B
364
0x016C
365
0x016D
366
0x016E
367
0x016F
368
0x0170
369
0x0171
370
0x0172
Size (By-
Data type Access
te)
1
U8
4
Float
1
U8
1
U8
1
U8
1
U8
1
U8
4
Float
4
Float
4
Float
4
Float
4
Float
1
U8
1
U8
4
Float
1
U8
19
String19
4
U32
4
U32
1
U8
1
U8
1
U8
1
U8
1
U8
1
U8
1
U8
1
U8
1
U8
1
U8
1
U8
Value range
R/W
0 = No, 1 = Yes
R/W
0 ... 15000
R/W
0 ... 10
R/W
0 ... 10
R/W
0 = No, 1 = Yes
R/W
0 ... 255
R/W
0 = distance,
1 = filling height
R
-
R
-
R
-
R
-
R
-
R
-
R
-
R
-
R
-
R
-
R
-
R
-
R
-
R
-
R
-
R
-
R
-
R
-
R
-
R
-
R/W
0 = Off, 1 = On
R/W
0 = Off, 1 = On
R/W
0 = Off, 1 = On
13 Bijlage