D e t e c t i o n
D e t e c t i o n
11. Specificaties
Size
Weight
Sensor technology
Temperature
Humidity
Alarm type
Alarm signal
Display
Direct Readout
Calibration
Datalogging
Capacity
Battery / Run time
Measurement
Housing
Sampling Pump
Charging and
communication
Response time T
90
Accuracy
Wireless range
Correction Factors
IP-Rating
EMI/RFI
Safety certifications
Warranty
NEO Foto-ionisatiedetector
260 x 78 x 58 mm (with boot)
708 g (24.9 oz) (with boot)
Photo-ionization sensor with standard 10.6 eV lamp (9.8 eV lamp in MP186)*
-20° to 50°C (-4° to 122°F)
5% ~ 95% RH (Non-condensing)
High, Low, TWA and STEL alarms
Over range alarm, battery low alarm
Man-Down alarm with pre-alarm and real-time re-mote wireless notification
Low Flow Alarm: Auto pump shutoff at low-flow condition
Acoustic: 95 dB @ 30 cm
Visual: flashing bright red LEDs, and on-screen indication of alarm conditions plus
wireless remote alarm notification
Vibration alarm
128 x 128 graphical LCD, 45 x 44 mm, with LED backlight for enhanced
display readability
Real-time reading of gas concentration (ppb, ppm, mg/m
gas and correction factor, lamp on/off, Man-Down alarm on/ off, battery status, pump
status, datalogging on/off, wireless on/off, temperature and time
Two/three-point calibration
Standard 12 months at one-minute intervals Storage interval adjustable from 1 to
3,600 seconds
Rechargeable Lithium-Ion battery with 24 hours typical operation
Pumped
Durable rubber boot, color coded for different models
Built-in pump with 2 settings from 300 to 430 cc/min Sample from up to 30 m (100 ft)
Charging, data download, instrument setup and firm-ware upgrades on PC or laptop
via Micro USB Wireless data and alarm status transmission via built-in RF modem
3 seconds (t
) VOC Mode
90
45 seconden @ 20°C (68°F) Benzene Tube Mode
±3% (at calibration point)
1,000 ft (300 m) line of sight
Integrated Correction Factor list of more than 200 compounds
IP67, IP66 when turned on
Highly resistant to EMI/RFI
Compliant with EMC Directive 2014/30/EU
UL/cUL: Class I, Div 1, Group ABCD
IECEx: Ex ia IIC T4 Ga
ATEX: II 1G Ex ia IIC T4 Ga
CE: Conformité Européenne
2 Years including lamp and sensor (1-year for 9.8 eV lamp)
Gebruikershandleiding
30
, μg/m
, PID measurement
3
3)