Harmonics IEC 61000-3-2
Voltage fluctuations
and flicker IEC 61000-3-3
EMC Immunity
Basic EMC
Phenomenon
standard
ELECTROSTATIC
DISCHARGE
61000-4-2
(ESD)
Radiated RF EM
Fields
61000-4-3
Proximity fields
from RF
wireless
com-
61000-4-3
munications
equipment
Electrical
fast
transients /
61000-4-4
bursts
Surges
Line-to-line
61000-4-5
Surges
Line-to-ground
61000-4-5
Conducted
disturbances
induced by RF
61000-4-6
fields
Bedieningshandleiding NeuroLight
Class B
Class A
Compliant
Professional
healthcare
facility
environment
Immunity Test
Levels
± 8 kV contact
IEC
± 2 kV, ± 4 kV,
± 8 kV,
± 15 kV air
3 V/m
80 MHz – 2,7
IEC
GHz
80 % AM at
1 kHz
Complies to
IEC
table 9 of IEC
60601-1-2
(2014)
± 2 kV
IEC
100 kHz
repetition
frequency
IEC
± 0,5 kV, ± 1 kV
± 0,5 kV, ±
IEC
1 kV,
± 2 kV
3 V
0,15 MHz –
80 MHz
6 V in ISM
IEC
bands
between 0,15
MHz and
80 MHz
80 % AM at
1 kHz
The NeuroLight must be used in professional
healthcare facility environment
The NeuroLight can be connected to the public
mains network
Compliance
levels
± 8 kV contact
In order to reduce ESD, the device
must be used in a 35% humidity
± 2 kV, ± 4 kV,
environment or more
± 8 kV air
Use of this equipment adjacent to or
stacked with other equipment should
3 V/m
be avoided because it could result
80 MHz – 2,7
in improper operation. If such use is
GHz
necessary, this equipment and the
80 % AM at
other equipment should be observed
1 kHz
to verify that they are operating
normally
Complies to
In order to prevent electromagnetic
table 9 of IEC
disturbance,
60601-1-2
separation from RF communication
(2014)
equipment of 30cm
The NeuroLight may temporarily not
± 2 kV
display
100 kHz
electromagnetic disturbances such
repetition
as the use of electrosurgery device. In
frequency
that case, the NeuroLight will maintain
the safety of the patient and the user.
Mains power quality should be that of
± 0,5 kV, ± 1 kV
a typical residential, commercial or
hospital environment.
Mains power quality should be that of
± 0,5 kV, ± 1
a typical residential, commercial or
kV, ± 2 kV
hospital environment.
3 V
0,15 MHz –
80 MHz
6 V in ISM
In order to prevent electromagnetic
bands
disturbance,
between 0,15
separation from RF communication
MHz and
equipment of 30cm
80 MHz
80 % AM at
1 kHz
EMC
Instructions/precautions
keep
result
during
keep
minimum
transient
minimum
21