6.Handling Errors and Problems
6.2

Troubleshooting

Problem
Arm cuff not applied
correctly.
Movement or talking
The reading is
during measurement.
extremely low (or high).
Clothing is interfering
with the arm cuff.
Is the air jack securely
inserted into the main
unit?
Arm cuff pressure does
not rise.
Air is leaking from the
arm cuff.
Arm cuff deflates too
The arm cuff is loose.
soon.
Is the cuff wrapped
correctly around the
arm?
Cannot measure or
readings are too high.
The cuff pressure does
not rise although the
pump motor can be
heard.
The unit loses power
The batteries are empty.
during measurement.
34
Cause
Remedy
Apply the arm cuff
correctly.
Refer to section 3.2.
Remain still and do not
talk during
measurement.
Refer to section 3.3.
Remove any clothing
interfering with the arm
cuff.
Refer to section 3.1.
Make sure that the air
jack is connected
securely.
Refer to section 3.2.
Replace the arm cuff
with a new one.
Refer to Chapter 8.
Apply the cuff correctly
so that it is firmly
wrapped around the
arm.
Refer to section 3.2.
Attach the arm cuff
correctly.
Check that the air plug
is properly connected to
the unit. Push the air
plug firmly into the air
jack.
Replace the batteries
with new ones.