Advertenties

OPERATION
 The meter is supplied complete with a 9V battery.
Slide off the battery compartment cover on the back
of the meter. Install the battery into the battery clip
connector while observing polarity.
 Always remove the electrode protective cap before
taking any measurement. If the electrode has been
left dry, soak the tip (bottom 2.5 cm) in rinse solution
(M10000B) for a few minutes to reactivate it.
 Connect the pH electrode to the BNC socket on the
top of the meter.
 Make sure that the meter has been calibrated
before taking any measurements (see Calibration
Procedure).
 Immerse the tip (2.5 cm) of the pH electrode into the
sample.
 Turn the instrument on by pressing the ON/OFF key.
 Allow the reading to stabilize before taking
measurements.
 After measurements, switch the meter off and store
the electrode with a few drops of storage solution
(SE-220) in the protective cap.
CALIBRATION PROCEDURE
 Remove the protective cap from the electrode.
 Rinse tip off with distilled water or a small quantity
of pH 7.
 Immerse the electrode tip into a new sachet of pH7
calibration solution and allow the reading to stabilize.
 Adjust the pH7 (OFFSET) calibration trimmer to display
"7.0 pH".
 Open a new sachet of pH4 calibration solution and
use a small quantity to rinse the electrode.
 Immerse the electrode tip inside the pH4 sachet and
allow the reading to stabilize.
 Adjust the pH4/pH10 (SLOPE) calibration trimmer to
read "4.0 pH".
 The calibration is now complete and the meter is
ready to take measurements.
 It is suggested the meter be calibrated at least once a
month, after a prolonged period of inactivity, and after
electrode replacement.
BATTERY REPLACEMENT
When the battery becomes weak the meter will
display "
".
When the low battery indicator appears, the battery has
only about 50 hours of working time left. A low battery will
result in unreliable measurements.
Prompt battery replacement is required. Battery
replacement must only take place in a non-hazardous
area using an alkaline 9V battery.
Turn the meter off, slide the battery compartment
cover located at the rear of the meter off and replace
the 9V battery with a new one. Make sure the battery
contacts are fully engaged in the connector, seat the
battery in its compartment and replace the cover.

OPTIONAL ACCESSORIES

pH 4.01 buffer solution,
M10004B
20 mL sachet (25 pcs)
pH 7.01 buffer solution,
M10007B
20 mL sachet (25 pcs)
pH 10.01 buffer solution,
M10010B
20 mL sachet (25 pcs)
Electrode rinse solution,
M10000B
20 mL sachet (25 pcs)
Electrode storage solution,
MA9015
220 mL bottle
MA950
Portable meter wall mounting kit
pH electrode with BNC connector and
SE-220
1 m cable