15. Technical results
EMC requirements
EN
IEC 60601-1-2 Edition 4.0 (2014)
Environment of intended uses:
Home Healthcare Facility Environment
Summary of Test Results
Emission (IEC 60601-1-2 section 7)
Radiated emission
Freq. range: 30 - 1000
MHz
Immunity (IEC 60601-1-2 section 8)
Immunity from
Electrostatic discharge
(ESD)
Immunity from radiated
electromagnetic fields
Immunity from Proximity
field from wireless
communications
equipment
Immunity from power
frequency magnetic field
Test
Standard
CISPR 11
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-3
IEC 61000-4-8
Class/ Severity level
Group 1 Class B
8 kV contact discharges &
15 kV air discharges
3.0; 10.0 V/m; 80 MHz
GHz,
80% AM, 1 kHz
List of frequencies,
from 9 V/m up to 28 V/m,
PM (18 Hz or 217 Hz), FM 1 kHz
30 A/m @ 50 Hz
26
Test
result
Complies
Complies
Complies
2.7
Complies
Complies