PART C – OPERATION
SECTION C1: Loading the trophon NanoNebulant Disinfectant Cartridge
The cartridge door opens automatically, DO NOT force it open.
Check expiry date of the trophon Nanonebulant, if expired it cannot be used to run
disinfection cycles.
A trophon NanoNebulant disinfectant cartridge needs to be inserted into the trophon2 device before a
HLD cycle can commence.
Refer to the trophon NanoNebulant IFU for detailed instructions on how to scan and log, insert, or
remove disinfectant cartridges.
Inserting a Disinfectant Cartridge
The screen on the trophon2 device will automatically prompt you to scan (if AcuTrace is enabled) and
insert a new disinfectant cartridge, when required. Follow the onscreen instructions and refer to the
trophon NanoNebulant IFU.
Note: If you are running trophon2 Software Version 1.5 and later, the trophon2 device will notify the
user upon scanning, if the trophon NanoNebulant cartridge is 30 days or less from expiration
be reminded again on the first cycle, 1 day before expiry.
The trophon NanoNebulant will also expire 30 days after insertion into the trophon2 device, regardless
of the expiry date shown on the disinfectant label.
Once trophon NanoNebulant is expired a purge will be required. Refer to SECTION E2: Purge Cycle.
SECTION C2: Logging the trophon Chemical Indicators
No other chemical indicators are approved for use in the trophon2 device.
Check expiry date of the trophon Chemical Indicators. Expired Chemical Indicators cannot
be used to run disinfection cycles.
If AcuTrace is enabled, log the new batch of trophon Chemical Indicators:
1.
Select Menu
2.
Follow the onscreen instructions.
Note: Chemical Indicators must be scanned at the start of every new carton. As this is a manual
process, it is important for users of trophon2 to complete this step each time a new Chemical Indicator
carton is opened. trophon Chemical Indicators must be stored in their original packaging and not be
shared across trophon devices outside of its original packaging.
Note: If you are running trophon2 Software Version 1.4 and later, trophon2 will notify the user during
the first cycle of the day if the trophon Chemical Indicators logged in the system are 30 days from expiry.
Users will be reminded again 5 days before expiry and everyday thereafter.
These reminders will not occur on earlier software releases. Do not use Chemical
Indicators after expiration date. Confirm Chemical Indicators are within expiry before
use.
If you are running trophon2 Software Version 1.4 and later, disinfection cycles will not
run after the Chemical Indicators have expired. Log trophon Chemical Indicators with
valid expiry date to continue use of the trophon2 device.
AcuTrace
Log Chemical Indicators.