If a cyclical recording attempt is made with the device running, a query of whether
the recording should be stopped automatically appears.
The device can only be switched off if the recording is stopped.
The Auto Power Off function is deactivated when a recording is in progress!
c) Delete logger recording:
Press and hold for two seconds: Start menu "
Switch the selection:
Switch the selection:
Confirmation of the selection, end of the delete menu

11. Universal output

The output can be used either as a serial interface (for USB 300 interface adapter) or as an
analogue output (0-1V). If the output is not required, it should be deactivated (Out oFF),
because the battery consumption is sharply reduced as a result.
If the device is operated with the USB 300 universal interface adapter, the device is supplied
from this interface.
Plug assignment:
 

11.1 Interface

With a USB 300 galvanically isolated interface converter (accessory), the device can be con-
nected directly to a USB interface of a PC. The transmission takes place coded in binary format
and is protected against transmission errors through extensive safety mechanisms (CRC).
The following standard software package is available:
• GSOFT3050: Operating and evaluation software for devices with the integrated logger
function
50
or
:
or
: Confirm with
Delete nothing (cancel the process)
Delete all data sets
Delete the most recently saved data set
4: external supply +5V, 50mA
3: GND
2:TxD/RxD (3.3V logic)
1: +UDAC, analog output
" appears.
: Confirm with
.
Only suitable
Adapter cable
are permitted
(Accessories)
.