8.3 Elektromagnetische emissie

This produt is intended for use in the electromagnetic environment specified below.
The customer or the user of this product should assure that the product is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/ flicker emissions
IEC 61000-3-3

8.4 Elektromagnetische immuniteit

Een onbedoelde beweging door de geteste eenheid is niet toegestaan voor EMC-goedkeuringscriteria.
This produt is intended for use in the electromagnetic environment specified below.
The customer or the user of this product should assure that the product is used in such an environment.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast transient
/ Burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
LikoGuard™ plafondlift • 7NL121101 Rev. 4
Guidance and manufacturer's declaration – electromagnetic emissions
Guidance and manufacturer's declaration – electromagnetic immunity
IEC 60601 test level
+/- 6 kV contact
+/- 8 kV air
+/- 2 kV for power
supply lines
+/- 1 kV for input/out-
put lines
+/- 1 kV differential
mode
+/- 2 kV common mode
<5 % U
T
(>95 % dip in U
)
T
for 0,5 cycle
40 % U
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
<5 % U
T
(>95 % dip in U
))
T
for 5 sec
3 A/m
Compliance
Electromagnetic environment - guidance
Group 1
The use of RF energy is only for its internal
function. Therefore, its RF emissions are
very low and are not likely to cause any
interference in nearby electronic equipment.
Class B
Suitable for use in all establishments other than domestic
and those directly connected to the
public low-voltage power supply network that
Class A
supplies buildings used for domestic purposes.
Complies
Compliance level
+/- 6 kV contact
+/- 8 kV air
+/- 2 kV for power sup-
ply lines
n/a. for input/output
lines
+/- 1 kV differential
mode
n/a. for common mode
3 A/m
25
Electromagnetic environment -
guidance
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environment