GROUND LEAKAGE
SHORT-CIRCUIT
OVERLOAD
CURRENT<1mA
CALL SERVICE
• You could also try a general RESET: switch the instrument on while
holding SET pressed. Attention! All stored programs and data will be
erased!
1. Select SET-UP and press SET.
2. Select DATA and press SET.
3. Select DURATION and press SET.
4. Select the desired time for the data-logging and press SET.
5. Select LOG and press SET.
6. Decide whether or not all data (voltage, current, power, time) during
the next experiments should be logged and press SET.
7. Select SET-UP and press SET to return.
• Starting a new experiment will erase the previous data in memory.
• Stored data can be transferred to a computer via the RS232 output
using optionally available software.
• At any time, you can return in the stand-by mode by holding SET
pressed during 2 seconds.
EV243
= A dangerous ground leakage is present (check
thoroughly the complete assembly).
= Connected tank is short-circuited or has a very
low resistance (check thoroughly the complete
assembly).
= Connected tank has a too low resistance (check
specifications).
= The output current is lower than 1 mA (switch
the alarm off in the SET-UP menu).
= Failure in the electronics.
***
ERROR
GROUND LEAKAGE
***
ERROR
SHORT-CIRCUIT
***
ERROR
OVERLOAD
***
ERROR
CURRENT<1mA
***
ERROR
CALL SERVICE
Data-logging
----
STAND-BY
↕
SET-UP
----
SET-UP
↕
DATA
-----
DATA
↕
DURATION
LOG DATA
FOR:
-----
DATA
↕
LOG
LOG
DATA:
-----
DATA
↕
SET-UP
Alarms
***
***
***
***
***
----
----
-----
↕
10h
-----
↕
YES
-----
10 /32