Technical Information

RS 232/RS 485 Communication Description

Window-Meanings

Tab. 7
N.
000
001
008
100
101
102
103
104
105
106
107
266/294
All manuals and user guides at all-guides.com
Read/Write
Datatype
R/W
L
R/W
L
R/W
L
R/W
L
R/W
N
R/W
N
R/W
N
R/W
L
R/W
N
R/W
L
R/W
L
TV 401/301/21 NSF User Manual / 87-900-995-01 (C)
Description
Start/Stop (in remote mode
the window is a read only)
Low Speed Activation
Remote (default) or
Serial configuration
Soft Start
(write only in Stop condition)
Set Point type
Set Point threshold
(expressed in Hz, mA or s)
Set Point delay: time between
the pump start and the set
point check (seconds)
Set Point signal activation
type: the signal can be "high
level active" or "low level
active"
Set point hysteresis
(in % of threshold)
Water cooling
Active Stop
(write only in stop)
Admitted Values
Start = 1
Stop = 0
NO = 0
YES = 1
(default = 0)
Remote = 1
Serial = 0
(default = 1)
YES = 1
NO = 0
0 = Frequency
1 = Current
2 = Time
(default = 0)
(default = 980)
0 to 99999
(default = 0)
0 = high level
active
1 = low level
active
(default = 0)
0 to 100
(default = 2)
0 = NO
1 = YES
0 = NO
1 = YES