5. During Use

Risk of electric shock during operation. Be very careful when
measuring live circuits.
If the equipment is used in a manner not specified by the manufacturer,
the protection provided by the equipment may be impaired.
Never exceed the limit value for protection. This limit value is listed
separately in the specifications for each range of measurement.
Do not touch unused terminals when the meter is linked to a circuit
which is being tested.
Never use the meter with CAT III installations when measuring voltages
that might exceed the safety margin of 600 V above earth ground.
Set the range selector at its highest position if the intensity of the
charge to be measured is unknown beforehand.
Disconnect the test leads from the tested circuit before rotating the
range selector in order to change functions.
When carrying out measurements on a TV set or switching power
circuits, always remember that the meter may be damaged by any high
amplitude voltage pulses at test points.
Always be careful when working with voltages above 60 VDC or 30 VAC
rms. Keep your fingers behind the probe barriers at all times during
measurement.
Never perform resistance, diode or continuity measurements on live
circuits. Make sure all capacitors in the circuit are depleted.

6. General Description

Refer to the illustration on page 2 of this manual:
1.
Display
3 ½ digits, 7 segments, LCD: 54 x 27 mm
2.
Hold button
3.
Backlight button
4.
Function button
"VmA" jack
5.
Insert the red (positive) test lead in this connector to measure voltage,
resistance and current (except 10 A).
"COM" jack
6.
Insert the black (negative) test lead.
"10A" jack
7.
Insert the red test lead in this connector in order to measure a max.
current of 10 A.