11. Reactivation of temperature monitor

In order to avoid an overheating and
consequently the destruction of impor-
tant Polymax components, all units
have been equipped with a tempera-
ture control (11). This part is situated
from outside at the bottom of the poly-
merisation pot and interrupts the cir-
cuit of the heating in case of too much
temperature development. In order to
reactivate the temperature monitor dis-
connect the unit from the mains and
let it cool down. Drain the water com-
pletely through the drain tap (23). Lay

12. Additional fields of application

All Polymax units are suitable for
pressure compression of embedding
masses, plasters and silicones. For
this purpose drain water through the
drain tap (23), close the tap again and
adjust the temperature to 0 °C (32 °F)
(see 7.1).
Important: Heating up the poly-
merisation pot without water
may lead to damage of the heating
cartridge.
44
the Polymax on its side and use a soft
tissue as underlay, if necessary.
Unscrew the perforated metal plate at
the bottom of the unit and push the red
reset knob of the temperature monitor
until it latches completely. Insert per-
forated metal plate again and tighten
it. Put the Polymax into normal posi-
tion, close drain tap (23), fill the unit
with water and connect to the mains
by means of the power cord. Then go
on as described from point 7 on.
Open the lid (2) at the handle (1) and
put the objects into the dry polymeri-
sation pot.
Tip: In order to avoid impurities and/
or unevenness at the bottom of the
pot caused by plaster or embedding
masses we recommend an underlay
of paper or the like.
Close the lid of the pot up to the block
and start pressure-polymerisation
(see 7.3).

13. Care and maintenance

Disconnect the unit from mains sup-
Attention: Do not use scouring
ply before care and maintenance! The
agents, steel-wool or sponge
water inside the pressure pot has to
scourers.
be changed regularly (at least once a
week), because the lime particulates
Clean the interior pot with fresh water,
harm the surface of the pressure pot.
a neutral washing-up liquid and a soft
scarf or sponge. Afterwards let the
Polymax 1, 3 and 5
water run out. Close the drain tap (23)
For the cleaning of the pressure pot
and fill with fresh water. The O-ring
open the drain tap (23) at the rear of
should be greased regularly with sili-
the unit and let the water flow out.
cone grease (35 g tube,
Remove dirt thoroughly.
D3350).