Table of Contents
1 Specifications
on page 3
2 Online user manual
on page 6
3 General information
on page 6
Section 1 Specifications
Specifications are subject to change without notice.
The product has only the approvals listed and the registrations, certificates and declarations officially
provided with the product. The usage of this product in an application for which it is not permitted is
not approved by the manufacturer.
1.1 Sensor specifications
Table 1 ORBISPHERE family of GA2X00 oxygen sensors
Specification
Type
Dimensions (Ø × L)
Weight
Pressure resistence
Materials
Other
Certification
Note: Unlike the GA2400 non-ATEX sensors, the intrinsically safe GA2800 ATEX sensors have no smart capability
to store calibration data. However, all GA2800 ATEX sensors have the ATEX conformity information engraved on
the sensor itself.
1.2 Sensor configurations
Application
Beer with ORBISPHERE
3650 Portable
Beer in-line
Wort in-line
Wort with ORBISPHERE
3650 Portable and special flow
cell 32007W.xxx
1
Viton and Kalrez are trademarks of DuPont Corporation.
Details—Non-ATEX sensors
40 bar maximum with default PPS collar (100 bar with stainless steel collar)
Stainless steel
®
Sensor head: Kalrez
O-rings
®
Sensor bottom: Viton
O-rings
Smart capability
CE
Table 2 Beverage
Sensor
GA2400-S00
GA2400-S00
GA2400-S00
GA2400-S00
4 Installation
on page 8
5 Maintenance
on page 13
6 Troubleshooting
Details—ATEX sensors
Electrochemical oxygen sensor
32 × 86.3 mm (1.26 × 3.40 in.)
300 g
Stainless steel or hastelloy
EPDM and Kalrez
Sensor head: Kalrez
1
Sensor bottom: Viton
Intrinsically safe
CE, Ex II 1 G, Ex ia IIC T6
Membrane Cartridge
2958A-A (optimized response
time)
2952A-A (optimized
maintenance interval)
2952A-A
29552A-A
29552A-A
on page 21
®
O-rings
®
O-rings
®
®
O-rings or Kalrez
O-rings
Protection Cap
33051-SP
33051-SG
33051-S0
33051-S0
English 3