5.2 Detector - Functies L
For electrical calibration, use the ADM0P05 adapter.
Function Duration of the tone burst (ms)
L
F
L
S
L
I
Maximum error for signals with crest factor ≤ 3:
Maximum error for signals with crest factor ≤ 5:
Maximum error for signals with crest factor ≤ 10:
Maximum overshot:
L
:
F
L
:
S
Maximum error of level linearity (31.5 to 12500 Hz):
Maximum error of differential level linearity( 31.5 to 12500
Hz):
Peak detector - L
Decay rate:
Onset time constant:
5.3 Peak detector - L
Onset time constant:
5.4 Integrator - L
Linearity range:
Pulse range:
Response time to a steady input signal:
, L
en L
F
S
Maximum error in the maximum response to a tone burst
200
500
20
5
function
I
functie
peak
, L
and L
functie
T
t
E
I
Maximum error (dB)
36
+ 1.0, - 2.0
± 2.0
± 2.0
± 3.0
± 1.0 dB
± 1.0 dB
± 1.5 dB
1.1 dB
1.6 dB
± 1.0 dB
± 0.3 dB
2.9 dB/s ± 1.0 dB/s
< 3.5 ms
< 75 μs
110 dB
65 dB
2 s