IMMUNITY LIMITS & MEASUREMENT
PHENOMENON
ELECTROSTATIC
DISCHARGE
RF
ELECTROMAGNETIC
FIELDS
ELECTRICAL FAST
TRANSIENTS
COMMON MODE /
BURSTS
SURGES LINE
TO LINE
SURGES LINE
TO GROUND
RF COMMON MODE
VOLTAGE DIPS &
INTERRUPTIONS
TEST
STANDARD
METHOD
IEC 61000-
IEC 60601-1-2
4-2
IEC 61000-
IEC 60601-1-2
4-3
IEC 61000-
IEC 60601-1-2
4-4
IEC 61000-
IEC 60601-1-2
4-5
IEC 61000-
IEC 60601-1-2
4-5
IEC 61000-
IEC 60601-1-2
4-6
IEC 61000-
IEC 60601-1-2
4-11
TEST LEVELS
±2 kV, ±4 kV, ± 6 kV, ± 8
kV - Contact Discharge
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV - Air Discharge
10 V/m
80 – 2700 MHz
80 % AM at 1 kHz
± 2 kV for A.C. Power Lines
100 kHz Burst Frequency
± 0.5 kV & ±1 kV At 0°,
90°, 180°, 270° ; Repetition
Rate: 60 S
± 0.5 kV, ±1 & ±2 kV At 0°,
90°, 180°, 270°
3 & 6 V for A.C. Main Port
; 80%, 1 kHz Amplitude
Modulation)
Frequency: 150kHz to
80MHz
0 % UT; 0,5 Cycle At 0°,
45°,90°, 135°,180°, 225°,
270°, 315°,
0 % UT; 1.0 Cycle
70 % UT; 25/30 Cycles
RESILT
PASS
PASS
PASS
PASS
PASS
PASS
PASS
163