18. EMC - Test Results
The device does not emit any radiation intended for
medical purposes, neither hazardous nor potentially
hazardous radiation.
PHENOMENON
RADIATED RF
EMISSION
CONDUCTED
RF EMISSION
VOLTAGE
FLUCTUATION
& FLICKER
162
EMISSION LIMITS & MEASUREMENT
STANDARD
TEST METHOD
CISPR 11
IEC 60601-1-2
ETSI EN 301
489-1
CISPR 11
IEC 60601-1-2
ETSI EN 301
489-1
IEC 60601-1-2
IEC 61000-3-3
TEST LEVELS
30~1000 MHz ; 1~3 GHz
& 3~6 GHz
(Measurement Uncertainty
4.24 dB)
0,15~30 MHz
(Measurement Uncertainty
3.35 dB)
EUT VALUES & LIMITS
Pst: 0.028 ; Limit 1.00
Plt: 0.028 ; Limit 0.65
dc [%]: 0.0000 ; Limit 3.30
dmax [%]: 0.053 ; Limit 4.00
dt [s]: 0.000 ; Limit 0.50
RESILT
PASS
PASS
PASS