Fault
Little or no aerosol is escaping.
The medication cap cannot be
removed from the medication reservoir
after the inhalation treatment.
If the malfunction persists, contact your local service partner.
®
eFlow
rapid - 01/17
Possible cause/Remedy
The ambient temperature should be at least
10 °C.
• Wait until the device has stabilised at the
lowest ambient temperature.
The negative pressure in the aerosol cham-
ber is too great.
• Open the aerosol chamber and gently pull
the aerosol head so that air is able to flow
in.
The cap can now be removed easily.
en
47