Section 2 Installation
®
Brooks
Ar-Mite
2-5 Transit Precautions
2-6 Removal From Storage
2-7 Installation of Flowmeter
2-8 Installation of Inductive Alarm
2-2
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
TM
MT3750C
To safeguard against damage during transit, transport the instrument to the
installation site in the same container used for transportation from the
factory if circumstances permit.
Upon removal of the instrument from storage, a visual inspection should be
conducted to verify its "as-received" condition. If the instrument has been
subject to storage conditions in excess of those recommended (See
Section 2-3), it should be subjected to a pneumatic pressure test in
accordance with applicable vessel codes.
Recommended installation for Model MT 3750C is as follows:
A. Carefully remove the covers from each end of the flowmeter.
B. Install the flowmeter with the inlet at the bottom and the outlet at the top.
C. When installing the flowmeter in the process line, follow accepted
plumbing practices for flanged or threaded fittings.
D. Install the flowmeter within 5° of true vertical. Use of a level is
recommended to determine the proper alignment.
E. Installation of a bypass piping arrangement is recommended, Refer to
Figure 2-1. Bypass piping permits the meter to be isolated from the flow
line for servicing and cleaning.
GENERAL NOTE FOR ELECTRONICS:
The Electrical connections shall be made in such a way that the degree of
ingress protection is maintaintained and is suitable for the installed
environment. See UL 50, NEMA 250 and EN 60529. Most installations will
require Type 4X or IP54 as a minimum installation.
Installation and Operation Manual
X-VA-MT3750C-eng
Part Number: 541B063AAG