Low Priority Alarms

The following low priority alarm messages are accompanied by a double beep
and a solid yellow light.
Indicator
Condition
4
Low Oxygen
3
Service Needed
2
Sensor Fail
Low Flow
Flow setting flashing

Medium Priority Alarms

The following medium priority alarm messages are accompanied by a triple beep,
repeated every 25 seconds, and a flashing yellow light.
Indicator
Condition
5
Oxygen Error
4
Flow Error
Action/ Explanation
Concentrator is producing oxygen at a slightly
decreased level. Contact your equipment
provider to arrange service.
The concentrator is operating to specification
but requires service at the earliest convenience.
Contact your equipment provider to arrange
service.
One of the concentrator's sensors has
malfunctioned. If the condition persists, contact
your equipment provider to arrange service.
The concentrator is producing oxygen at a
slightly decreased rate. If condition persists,
contact your equipment provider to arrange
service.
Action/ Explanation
The concentrator is producing oxygen at a
low concentration. Switch to a backup oxygen
source and contact your equipment provider to
arrange service.
Proper oxygen delivery has not been detected.
Check for kinked tubing. If condition persists,
switch to a backup oxygen source and contact
your equipment provider to arrange service.