VisionAire
3
™
IV. 6. Technical characteristics
Dimensions: D x W x H: 29.2 x 35.8 x 52.8 cm (11.5 x 14.1 x 20.8 in).
Weight: 13.6 kg (30 lbs).
Sound level: 40 dBA
Flow values:
1 - 3 l/min ± 10% of indicated or ± 200 ml whichever is greater
from 0 - 7 kPa (0 -1 psig back pressure).
Average oxygen content:
• 90% +5.5%/-3% @ 1 - 3 l/min indicated
(values at 21°C (69.8ºF) and at an atmospheric pressure corresponding to 0 - 4000 m). Max. flow:
3l/min. The maximum outlet pressure is 55kPa.
Electrical power supply:
• 230 V - 50 Hz Europe / 115 V – 60 Hz / 230 V – 60 Hz (other countries,
depending on version)
• Use the cable provided, and check that the electrical characteristics of the mains socket used
match those indicated on the manufacturer's plate on the rear panel of the machine.
• Mean power rating: 175 watts
Class II
Type B
1.5 amps (230 V).
3.0 amps (115 V).
Filters:
Before the oxygen outlet: a product filter < 0.30 µm.
Air circulation:
A fan cools the compressor compartment.
Environmental limit conditions:
The performances of the device (especially the oxygen concentration) are quoted at 21°C (69.8ºF)
and 1013 mbar. They may change with temperature and altitude.
- The device should preferably be stored and transported in the vertical position.
- The device must only be used in the vertical position.
- Ambient temperature of between 5°C and 40ºC (41ºF and 104ºF) (operation for 115 VAC units).
- Ambient temperature of between 5°C and 35ºC (41ºF and 95ºF) (operation for 230 VAC units).
- Storage temperature range from -20°C to 60°C (-4ºF and 140ºF) .
- Relative humidity up to 95% (non-condensing).
AirSep
Corporation
®
MN157-1 Rev - 11/09
GB
F