Cleaning
Unless contamination is suspected, periodic cleaning is all that is recommended. The Airflo 24 should be
cleaned/disinfected as follows:
• Unplug the Airflo 24 from the battery charger. Clean the external surfaces using a non abrasive damp
cloth with warm water. Wipe dry thoroughly with a dry towel before reconnecting to the battery
charger. Do not allow moisture to enter any part of the compressor.
• Do not use solvent-based products.
Disinfecting/Sterilising
• Disinfection – Pre-clean the Airflo compressor by wiping with a damp cloth and then disinfect by
wiping with a 1:10 solution of household liquid bleach and water - 1 part bleach to 10 parts cold
water (10,000ppm (1%) Sodium Hypochlorite solution). Do not allow moisture to enter any part of
the compressor. Allow to air dry. In usual circumstances this product does not require sterilisation for
reuse.
Storage
• When not in use, store in a dry place together with these instructions. To maintain the batteries, when
being stored for long periods, leave the Airflo 24 on charge.

Maintenance

- fig.
The Airflo 24 has been designed to be maintenance free and
apart from replacing the battery, the Airflo contains no other user
serviceable parts.
Battery replacement:
• Turn the Airflo 24 onto its end containing the air outlet socket,
remove the screw (A) and slide the end panel (B) away from
the housing.
• Disconnect the wire from the battery pack and remove the
battery pack from the battery compartment.
• Fit the replacement battery pack, ensuring that the connector
is secure, refit the end panel and secure the screw (A).
• To calibrate the battery indicator, it is essential before use to
charge for 12 hours.
When it becomes apparent that the Airflo 24 is producing fewer than expected lifting cycles, the battery is
probably reaching the end of its working life and should be replaced. For optimal performance, Winncare
recommends replacing batteries every 12 months.
If a fault should develop or if you require assistance in setting up using or maintaining the equipment, please
contact your supplier.
Disposal
When the Airflo 24 has reached the end of it's working life, remove the battery for safe disposal in
accordance with local regulations, and dispose of the Airflo 24 at your local household waste recycling
centre (civic amenity site).