Storing
The Lode equipment is intended for use in the electromagnetic environment specified below. It is the
responsibility of the customer or user to ensure that the Lode equipment is used in such an
environment.
Immunity test: Conducted RF 61000-4-6
Immunity test:
IEC 60601-Test Level: 3 Vrms 150kHz to 80MHz
IEC 60601-Test Level:
Compliance Level: 3 Vrms
Compliance Level:
Immunity test: Radiated RF 61000-4-3
Immunity test:
IEC 60601-Test Level: 3 V/m 80MHz to 2,5 GHz
IEC 60601-Test Level:
Compliance Level: 3 V/m
Compliance Level:
Electromagnetic Environment - Guidance: Portable and mobile RF communications equipment should
Electromagnetic Environment - Guidance:
be used no closer to any part of the Lode equipment including cables, than the recommended
separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended seperation distance:
Recommended seperation distance:
d= 1,2 √ P 80 MHz to 800 MHz
d= 2,3 √ P 800 MHz to 2,5 GHz
where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey a , should
be less than the compliance level in each frequency range. b
Interference may occur in the vicinity of equipment
marked with the following symbol:
M960903 - MANUAL CORIVAL PEDIATRIC / VERSION: 1.2 - 2016-11-29
20/24