Download Inhoudsopgave Inhoud Print deze pagina
Inhoudsopgave

Advertenties

13.2 - Immunity test

The Lode equipment is intended for use in the electromagnetic environment specified below. It is the
responsibility of the customer or user to ensure that the Lode equipment is used in such an
environment.
Immunity Test: Electrostatic discharge (ESD) EN 61000-4-2
Immunity Test:
IEC 60601-Test Level: ± 6 kV contact
IEC 60601-Test Level:
± 8 kV air
Compliance Level: ± 6 kV contact
Compliance Level:
± 8 kV air
Electromagnetic Environment - Guidance: Floors should be wood, concrete or ceramic tile. If floors are
Electromagnetic Environment - Guidance:
covered with synthetic material, the relative humidity should be at least 30%.
Immunity Test: Electrical fast transient/burst EN 61000-4-4
Immunity Test:
IEC 60601-Test Level: Not applicable
IEC 60601-Test Level:
Compliance Level: Not applicable
Compliance Level:
Electromagnetic Environment - Guidance: Conducted only on ports interfacing with cables whose total
Electromagnetic Environment - Guidance:
length, according to the manufacturer's functional specification, may exceed 3 meters
Immunity Test: Surge EN 61000-4-5
Immunity Test:
IEC 60601-Test Level: Not applicable
IEC 60601-Test Level:
Compliance Level: Not applicable
Compliance Level:
Electromagnetic Environment - Guidance: Conducted only on ports interfacing with cables whose total
Electromagnetic Environment - Guidance:
length, according to the manufacturer's functional specification, may exceed 3 meters
Immunity Test: Voltage dips, short interruptions and voltage variations on power supply input lines EN
Immunity Test:
61000-4-11
IEC 60601-Test Level: Not applicable
IEC 60601-Test Level:
Compliance Level: Not applicable
Compliance Level:
Electromagnetic Environment - Guidance: Equipment for professional use with a total rated power
Electromagnetic Environment - Guidance:
greater than 1 kW and is excluded from this requirement
Immunity Test: Power frequency (50/60Hz) magnetic field EN 61000-4-8
Immunity Test:
IEC 60601-Test Level: 3 A/m
IEC 60601-Test Level:
Compliance Level: 3 A/m
Compliance Level:
Electromagnetic Environment - Guidance: Power frequency magnetic fields should be at levels
Electromagnetic Environment - Guidance:
characteristics of a typical location in a typical commercial or hospital environment.
NOTE: Ut is the AC mains voltage prior to application of the test level.
NOTE:
M960903 - MANUAL CORIVAL PEDIATRIC / VERSION: 1.2 - 2016-11-29
19/24

Advertenties

Inhoudsopgave
loading

Gerelateerde Producten voor Lode Corival 2015

Deze handleiding is ook geschikt voor:

M9609 series

Inhoudsopgave