GEBRUIKERSHANDLEIDING en TECHNISCHE HANDLEIDING KANMED-WARMHOUDKAST
7.2
TROUBLE SHOOTING
If the warming is not starting, check as follows:
Power in the wall socket?
Power Switch on (If power is OK it will show a green light)
Power cable connected (UK only - plug fuse OK)?
Cabinet fuses OK?
Is the fan rotating? It shall start as soon as the power switch is on. If it is not
rotating, making a strange noise or rotates slowly it must be changed.
Check the fan and that the air intake under the heating compartment is not blocked.
Red lamp lights up now and then. Check the capillary thermostat and the
temperature regulating unit.
NOTE: All checking that involves opening the heating compartment must be done by a
qualified technician. Ensure that the cables to the heating compartment do not get stuck
when you close the heating compartment after service actions
7.3
CHANGINGTHE MAXIMUM TEMPERATURE, UP OR DOWN.
Follow the procedure described in 7.1.2 and re-adjust T2 so that it is 4-5
T1.
NOTE: Think carefully about the risks and consequences of setting the max temp limit
higher than the factory set temperature and about how to make the staff aware that a
higher temperature might have been set by someone not authorised to change the
temperature.
7.4
ACOUSTIC HIGH TEMP ALARM AND TEMPERATURE LOCK
Please contact Kanmed if you want to activate an acoustic over temperature alarm or
lock the possibility to change temperature at all..