PROCEDURE
Place the external control sensor as described under GENERAL INFORMATION
above.
Start the warming and wait until the temperature is stable (at least 1 hour).
Compare the external control thermometer with the set temperature and if the
deviation is bigger than +/-1
Regulator type No 700-0821
1. Press
and
don't succeed- .repeat) and keep buttons pressed for 5 sec until the indicator
display shows "PA".
2. Press
one time.
3. Press
within 15s until -19 is displayed.
4. Press
one time.
5. Press
and
displayed
6. To change a parameter press
7. Press SET once and within 2 sec on
difference. Press SET again
For example: The display shows 50
47
C. The difference is -3
Store the new calibrated value by pressing
may be heard, if you don't succeed- repeat) and keep pressed for 5 sec until the
indicator display shows the actual temperature.
8. Make a new measurement of the temperature and control that T1 shown value
compares to that of the external thermometer.
7.1.2
OVER TEMPERATURE T2 CALIBRATION AND CHECK
The temperature sensor T2 is a capillary thermostat in series with the heating element.
In the Kanmed Cabinet its hysteretic is 4
the maximum value of T1. The activation of T2 is identified by a clear click sound as well
as when the red over temperature lamp is lit
The procedure described below is a simplified method to check and if necessary adjust T2
PROCEDURE
1. Follow the procedure above, step 1-5.
2. To change a parameter press
Normally set to 42˚C or 70˚C. Press "set" and change it to 47 and press set
again.
3. Exit the program by pressing the up and down buttons
view is present.
4. Press Set again and set the temp to 47˚C and let the cabinet temperature
stabilise.
5. When the temperature is stable adjust the over temperature by turning T2's
adjustment screw with a screwdriver (you find it through the hole on the
underneath of the shelf). Turn the adjustment screw fully clockwise (+). Now Turn
counter clockwise back to activation (one click sounds), turn again clockwise past
the activation point in tiny steps (a new click sounds). You have now set the over
temperature to about 47˚C.
6. Repeat the procedure 1-3 and set back the maximal selectable temperature to
42˚C.
7. Check that you don't activate the over temperature indication when running the
cabinet in normal operation (with T1 prepared on 42C) If this should happen you
have to repeat the procedure and set T2:s activation point a bit higher.
C adjust as follows.
exactly simultaneously (only one beep may be heard, if you
exactly simultaneously until "SP"(Select Parameter) is
or
or
C but the external control temperature shows
C. To compensate this difference press
C max. When delivered T2 is set to 5
or
button till parameter "CA1" is received.
button to enter the noted temperature
and
simultaneously (only one beep
button till parameter "r2" is received.
button 3 times.
C above
and
until normal