1. Specificaties

Vertikale deflectie:
Operating Modes:
Channel 1 or 2 only
Channels 1 and 2
(alternate or chopped)
Sum or Difference of
CH 1 and CH 2
Invert:
CH 2
XY Mode:
CH 1 (X) and CH 2 (Y)
Bandwidth (-3 dB):
DC, 5mV/div....20V/div.: 0...40MHz
AC, 5mV/div....20V/div.: 2Hz...40MHz
DC, 1...2mV/div.: 0...10MHz
AC, 1...2mV/div.: 2Hz...10MHz
Rise Time (calculated):
‹ 35 ns (1 mV/div....
Triggering:
Automatic:
Linking of peakdetection and triggerlevel
Min. signal height 0.5div
Frequency range 5Hz...50MHz
Level control range From peak- to peak+
Normal (without peak):
Min. signal height 0.5div
Frequency range 0...50MHz
Level control range –10....+10div.
Slope: Rising or falling
Sources: Channel 1 or 2, Line and External
Horizontale deflectie:
Time Base:
100 ns/ div.....0.2 s/div
(1-2-5 Sequence)
Accuracy:
± 3%
Variable (uncalibrated): › 2.5 :1 to › 1.25 s/div.
X Magnification x 10: up to 10 ns/div.
Operation/readout/control:
Manual:
via controls and buttons
Autoset:
automatic signal related parameter settings
Save and Recall: 6 instrument parameter settings
Component tester:
Test Voltage:
approx. 7 Vrms (open circuit)
Test Current:
max. 7mArms (short-circuit)
Test Frequency:
approx. 50 Hz
Test Connection:
2 banana jacks 4 mm O
One test circuit lead is grounded via protective earth (PE)
Algemeen:
CRT:
D14-363GY, 8 x 10 div. with internal
graticule
Acceleration Voltage:
approx. 2 kV
Trace Rotation:
adjustable on front panel
Z-Input (Intens. modulation):
max. + 5 V (TTL), 10 kHz
Probe ADJ Output:
1 kHz/1 MHz Square Wave Signal
ca. 0.2 Vpp
(tr ‹ 5 ns) for probe adjustment
Air-Parts BV
Deflection Coefficient:
Variable (uncalibrated):
Input Impedance:
Input Coupling:
Max. Input Voltage:
Coupling:
Trigger Indicator:
External Trigger:
Input Impedance: 1M II 15pF
External Trigger Signal: 0,3 Vpp ≤ 5 V,
DC (0...50MHz), AC (20 Hz...50 MHz)
Max. input voltage: 100V (DC + Peak AC)
Active TV sync. separator: Field and Line, +/-
Accuracy:
Hold-Off Time:
XY
Bandwidth X amplifier:
XY Phase shift
Power Supply (Mains):
Power Consumption:
Safety class:
Operating temperature:
Storage temperature:
Rel. humidity:
Dimensions (W x H x D): 285 x 125 x 380mm
Weight:
November 2010
2 mV/div.)
‹ 8,75 ns (5 mV/div....
20 V/div.)
1-2-5 Sequence
±5% (1mV/div....2mV/div.)
±3% (5mV/div....20V/div.)
› 2.5:1 to
› 50 V/div.
1 M II 15 pF
DC, AC, GND (ground)
400 V (DC + peak AC)
AC (5 Hz...80 MHz),
DC (0...80 MHz),
LF (0...1.5 kHz)
LED
± 5%
variable to approx. 10 : 1
0...2.5 MHz (-3 dB)
‹ 3°: ‹ 120 kHz
105...253 V, 50/60 Hz ±10 %,
CAT II
approx. 30Watt
at 230 V/50Hz
Safety class I (EN61010-1)
+5...+40° C
-20...+70°C
5...80% (non condensing)
approx. 4.8kg