14 EMC Information

Guidance and manufacturer's declaration – electromagnetic emmissions
The WarmCloud WarmCloud system is intended for use in the electromagnetic environment specified below. The customer or
the user of the system should assure that it is used in such an environment.
Emission test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Recommended separation distances between portable and mobile RF communications equipment and the
The WarmCloud WarmCloud system is intended for use in the electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the system can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters) and the WarmCloud
WarmCloud system as recommended below, according to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
150 kHz to 80 MHz
W
d = 1,17 √ P
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m)
can be established using the equation applicable to the frequency of the transmitter, where P is the maximum output power
rating in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interruptions
and voltage variations on the
power supply input lines
IEC 61000-4-11
Compliance
The WarmCloud WarmCloud system uses RF energy only for its internal
Group 1
function. Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
Class B
The WarmCloud WarmCloud is suitable for use in all establishments,
including domestic establishments and those directly connected to the
Class B
public low-voltage power supply network that supplies buildings used
for domestic purposes.
Complies
WarmCloud WarmCloud system.
Separation distance according to frequency of transmitter
80 MHz to 800
MHz
d = 1,17 √ P
0,2 m
0,2 m
0,4 m
0,4 m
1,2 m
1,2 m
3,7 m
3,7 m
11,7 m
11,7 m
IEC 60601 test level
±6 kV contact
±8 kV air
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV differential mode
±2 kV common mode
<5% U
T
(>95% dip in U
)
T
for 0,5 cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
Page 21 of 23
Gebruikershandleiding OT-600-070/5
Electromagnetic environment – guidance
m
800 MHz to 2,5 GHz
d = 2,33 √ P
0,3 m
1,6 m
2,3 m
7,4 m
23,3 m
Compliance level
±6 kV contact
±8 kV air
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV differential mode
±2 kV common mode
<5% U
T
(>95% dip in U
)
T
for 0,5 cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
Electromagnetic environment
– guidance
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%
Mains power quality should be
that of a typical commercial or
hospital environment
Mains power quality should be
that of a typical commercial or
hospital environment.
If continued operation during
power mains interruptions, it is
recommended that the Kanmed
WarmCloud system be powered
from an uninterruptible power
supply unit (UPS)