OPERATION
 Connect the supplied 12 VDC power adapter to the
controller and to the mains.
 Make sure the controller has been calibrated
before taking any measurements (see Calibration
Procedure).
 Immerse the tip (4 cm) of the Conductivity probe
into the sample.
 Turn the instrument on by pressing the ON/OFF key.
 Wait for thermal equilibrium to be reached and the
controller will start continuous monitoring.
 Set the BELLOW / ABOVE alarm activation switch in
the desired position.
 A blinking alarm will indicate when the measured
EC or TDS value is lower/higher than the setpoint
(in accordance the BELLOW / ABOVE switch).
Note: The output power contact has no protection
fuse inside the controller. It is recommended to
protected it outside, against failure.
SETPOINT
 The setpoint can be selected by adjusting the
central front knob to the desired value.
 The selectable range for MC311 is from 0.8 to
2.8 mS/cm.
CALIBRATION PROCEDURE
 Clean the probe with alcohol and let it dry (if a more
thorough cleaning is required, brush the metal pins
with fine sandpaper and then rinse with water).
 Open a sachet of M10031B (1413 µS/cm) conductivity
calibration solution and immerse the probe, making
sure that metal pins are completely submerged.
 Turn the controller on by pressing the ON/OFF key.
 Wait until the reading stabilize for temperature
variations, and then adjust the EC/TDS calibration
trimmer on the front panel with the supplied
screwdriver until the display shows: "1.4" mS/cm.
 The calibration is now complete and the controller
is ready for use.
 The instrument should be re-calibrated at least
once a month, or when the probe is changed.