Operation
• Fitting the battery (2 x AAA (1.5 V)
Change the batteries with this display
indication. Check position of the O-Ring
when opening/ closing the battery chamber lid!
 Turn the unit on using the ON/OFF key.
The display shows: "0". Make one Zero for all the
methods:
 Swirl the instrument several times under the
water surface allowing the sample chamber
to completely ill. Take the instrument out of the
water.
 Close the blue lid tightly.
 Press the ZERO/TEST key. "000" lashes for
approx. 8 seconds. The display shows "0.0.0".
 Select the test required using the MODE key:
Cl (free Chlorine/ total Chlorine) → pH → TA →
CyA → Cl ... (Scroll).
Open the blue lid and add the appropriate tablet
6
straight from the foil.
Method
Cl
: Free chlorine DPD No.1
Cl
: Total chlorine DPD No.1 + DPD No.3
pH : pH Value
CyA : Stabilizer
TA
: Alkalinity-M
{ Crush the tablet using a clean stirring rod.
Stir until the tablet is dissolved.
| Close the lid and swirl the instrument gently
for approx. 15 seconds.
} Afterwards keep the instrument steady and wait
for 15 seconds when testing ClF/pH/CyA and
2 minutes when testing Clt and TA.
4
Required Reagent(s)
PHENOL RED PHOTOMETER
CyA-Test
ALKA-M-PHOTOMETER
GB