We are pleased you have purchased our product and we thank you for your trust in us.
It is very important that the health care worker and/or the patient read and understand
INTENDED USE
Pulmowaves 2 is a medical device is for respiratory rehabilitation physiotherapy and in particular for the removal
of tracheobronchial secretions, including a nebulising device on demand.
The Medical Device should be prepared and used in the presence of or by qualified personnel for respiratory
rehabilitation physiotherapy who have read and understood this instruction manual provided, or adults who have
received specific information from the aforementioned qualified personnel and who have read and understood
this user manual provided.
The medical device can also be used for home-care treatment.
The therapeutic session may be set to (5-10-15-20 minutes) according to the doctor's assessment and the physical
condition of the patient.
The device does not require lubrication, is manageable, simple to use, reliable, resistant and quiet.
Note: Use only original FLAEM accessories
OPERATING INSTRUCTIONS
- All drugs can be nebulised with the device.
- COPD chronic obstructive pulmonary disease (Stage 0 to Stage IV in accordance with the GOLD guidelines).
- Bronchiectasis.
- Cystic Fibrosis.
- Neuromuscular, orthopedic or other disorders involving the respiratory system (bronchial hypersecretion,
ineffective cough) or who needs a quick effective removal of secretions.
CONTRAINDICATIONS FOR USE
- Children younger than 3 years old.
- Patients with cardiovascular instability.
DEVICE NOMENCLATURE
A
Respiratory physiotherapy device
A1 Switch
A2 Digital pressure gauge
A3 Flow regulator with filter
A4 Pulsed air output
A5 12Vdc output power supply socket for Nebulise
A6 Pressure switch input
A7 Filter tube
A8 Filter
A9 Pressure switch tube
A10 Pulsed air tube
A11 Power supply socket
A12 Power supply cable
A13 Tube connector clip
A14 Protective gasket
COMPOSITION
The device is a system composed of a pneumatic compressor, an ultrasonic nebulising device used to nebulise
saline solution only, and a dispenser equipped with a PEP (Positive Expiratory Pressure) device.
PulmoWaves 2
Device for respiratory rehabilitation physiotherapy
the device use and maintenance information.
B
Nebuliser Device
B1
Device switch
B2 Nebuliser
B2.1 Lower part
B2.2 Upper part
B2.3 – Support with valve
B3
Nebuliser power supply socket
B4
Power supply cable for nebuliser device
B5
Cable plug B4
C
ACCESSORIES
C1 PEP adjustment ring
C2 Mouthpiece
C3 PEP Fitting
C4 Nose clip
C5 Valve for use without nebuliser device
12