Check the con-
nection status
Selecting settings
Module status
Select the
Active
One of the following messages may appear in the display:
Active
-
The module is ready for operation.
Initialising, module not yet active
-
The module has not finished initialising. This process takes approx.
90 seconds.
Inactive
-
A module has not be recognised.
Check the module installation. Switch the washer-disinfector off
and then back on again.
If the message appears in the display again call Miele Service.
Document all settings in case of a subsequent Service call require-
ment.
For manual entries DHCP must first be deactivated.
DHCP
Select the
option and confirm your choice with OK.
DHCP
On
Off
Off
Select the
option and confirm your choice with OK.
The following settings are then available:
-
IP address
Subnet mask
-
Standard gateway
-
DNS server automatic
-
-
DNS server 1
DNS server 2
-
Port type
-
Port
-
Select an option and confirm your choice with OK. You can then
carry out settings as shown in the display.
en - Configuring the interface
option and confirm your choice with OK.