operATIng InsTruCTIons
CAUTION– For proper operation, ensure the flow generator is OFF
before inserting humidifier chamber.
3.
Slide the chamber back into the humidifier cradle. Ensure
the chamber is fully engaged with the humidifier cradle and
the chamber release latch snaps in place when inserted.
WARNING
Do not attempt to fill the humidifier while it is
attached to the humidifier cradle. Damage to the
humidifier cradle may occur.
If the system is placed on the floor, do not position it where it may be
stepped on when getting out of bed.
Never touch the heater plate on the humidifier cradle. Never touch the
heat transfer plate on the bottom of the water chamber. These plates can
reach temperatures as high as 149°F (65°C) during operation.
Do not operate the heater if the water chamber is empty. The heater
plate may be turned off using the keypad heater control when being used
without water.
OPERATING INSTRUCTIONS
Heater Controls
The heater controls and status are located on the keypad and
display.
1.
Heater Setting Control Buttons
2.
Heater Power LED
Controlling The Humidity
The heater setting can be adjusted in a range of 1 to 10. The
optimum temperature setting is dependent on the ambient
temperature and humidity of your room. If your room temperature is low and/or the
relative humidity is high, then a lower temperature setting may be necessary to prevent
excess condensation in the air supply tubing. If the room temperature is high and/or the
relative humidity is low, then a higher temperature setting may be necessary for optimum
comfort.
The heater setting of 1 results in a heater plate temperature of approximately 84°F (29°C).
The heater setting of 10 results in a heater plate temperature of approximately 149°F (65°C).